Q:

Which pipette tips are compatible with the 96-well Shuttle® Nucleocuvette® Plates?

A:

We recommend using special tips, e.g. epT.I.P.S™ (Eppendorf, Hamburg, Germany) or TallTips™ (Matrix Technologies, Hudson, NH, USA). You can contact our Scientific Support Team for a list of compatible tips, including tips for pipetting robots.

Categories:
Transfection

You may also want to know:

What microtiter plates would you recommend for running a MycoAlert® and MycoAlert® Plus Assay on a plate luminometer?
What is the difference between the MycoAlert® and MycoAlert® PLUS Assay?
Do the MycoAlert® and MycoAlert® PLUS Mycoplasma Detection Kits detect dead mycoplasma?
Is it possible to store reconstituted MycoAlert®/MycoAlert® PLUS Reagent and Substrate?
Do antibiotics effect the MycoAlert® and MycoAlert® PLUS Mycoplasma Detection Kits?
Are there positive controls for the MycoAlert® Assay?
Which luminometer is suited for performing the MycoAlert® Assay?
For hepatocytes, is plating density a concern post Nucleofection®?
Can I use the ViaLight™ Plus Assay also, if my cells are cultured in 24well plates, 6well plates or 12 well plates?
Is the MycoAlert® Assay validated for use in GMP environment?
See More

Privacy | Legal | About Lonza

© Lonza. All rights reserved.